2024年4月1日发(作者:)
GE Healthcare
Women’s Healthcare
Senographe
Pristina
GE imagination at work
The Senographe* Pristina is a full field digital mammography
system designed to offer an extensive breast care solution
with screening and diagnostic capabilities, focused on an
ergonomic design for the technologist and patient comfort.
Senographe Pristina features a 24 x 29 cm detector, designed
to offer full breast coverage in a single image. Smaller breasts
can also be imaged in any view with paddles that can slide to
both sides of the detector.
The Senographe Pristina does not require daily calibration.
Ergonomics for technologists
• Re-imagined user interface
• Park Positioning during patient positioning
• One touch access to preset rotation for positioning
• Variable speed motorized gantry movements
• Sliding compression paddles can move to the side of the
detector for compression
Ergonomics and design for patient
comfort
• Designed for Patient comfort
• Wheelchair access, MITA compliant
• Thinner Bucky than previous platform
• Rounded edges detector for patient comfort
Image quality
• Automatic Optimization of Parameters (AOP), selects all
exposure parameters based on breast radiological properties
• Three AOP modes + 1 Automatic mode for implants
• eContrast is an image processing feature that makes
automatic adjustments of brightness and contrast
• DQE at IEC 62220-2-3 equivalent spectrum, at 75µGy:
70% (+/-3) at 0.5lp/mm and 64% (+/-3) at 2lp/mm
Smooth digital workflow connectivity
• Automated Quality Control
• Integrated Repeat and Reject Analysis
Technical Specifications
Detector
• Detector ready to use right after system boot
• Detector size: 24 x 29 cm
• Pixel size (pitch): 100 μm
• Acquisition dynamic range: 14 bits
• Bucky front cover thickness: 49mm
• Optimized room for positioning due to the bucky depth:
470mm
• Image size:
– LFOV image size - approx. 13 MB per image
– Regular image size - approx. 9 MB per image
• Patented needle structure CsI scintillator, single piece
construction
• Breast support with rounded edge
• Air cooling
Tube technology
• X-Ray tube type: Artemis
• Anode target materials - Dual track: Molybdenum (Mo)
enriched with Vanadium, and Rhodium (Rh)
• Four focal spots: 0.1 and 0.3 IEC on each target
• Target angle: 0 degree
• Maximal high voltage: 49 kV
• Tube current:
– Molybdenum target:
• 100 mA from 25 to 30 kV on large focal spot
• 40 mA from 25 to 30 kV on small focal spot
– Rhodium target:
• 62 mA from 25 to 30 kV on large focal spot
• 35 mA from 25 to 30 kV on small focal spot
• Anode size (tracks diameter): 100 mm
• Anode heat storage capacity: 250kJ (340 kHU)
• Anode maximum dissipation: 500 W (40 kHU/min)
• Max casing continuous dissipation:
150 W (12 kHU/min) at 40 °C
• Permanent filtration: 0.69 mm Beryllium
• Weight: 7 kg
• X-ray tube assembly: self-encased X-ray tube, oil-free,
lead-free, air-cooled head
• Tube protection: software monitoring of tube load
Grid/breast support
•Universal grid compatible with 2D Conventional
Mammography and DBT
• Ergonomic breast support designed for patient comfort
and cleanability
• Motorized lock of the grid and breast support
• Breast support material: carbon fiber composite
• Optimized grid motion ensuring no grid structure visible in the
image
• Detector to breast support edge-to-edge distance ≤ 5 mm