332・ 中国煤炭工业医学杂志2015年2月第18卷第2期Chinese Journal of Coal Industry Medicine,Feb.201 5,Vo1.18,No.2

社会医学与卫生事业管理・

科学确立日立H7180全自动生化分析仪校准周期

王春雨 李宇周 张桂丽

摘要 目的 利用校准周期图科学确立日立7180全自动生化分析仪校准周期。方法 生化分析仪定标

校准后,在距定标校准时间为0,2,4,6,8,10,12,24h时分别测定高、中、低三种浓度水平的质控血清1

次,之后每24h测定1次,总计测定30d,如果有项目稳定性强可继续延长检测时间,出现偏差后结束。所

有检测项目均重复测定4次取平均值。利用SPSS 19.0统计软件进行数据处理,计算累积变异系数

(CV),并绘制校准周期图。从校准完毕后开始计时至CV>1/6 CI A'88允许误差时结束,此间隔时间即

为该项目的校准周期。结果 校准周期最短为ALP只有14h;钙(Ca)为24h;TP、AI B、CK、Fe为4~10d;

DBil、TBi1、HDL—C、GLU、Urea、CH0为10~20d;LDH为25d;TG去游离、Cr酶法、UA、AMY、

ALT、AsT的校准周期超过30d。结论 科学、合理地确立各检测项目的校准周期,对生化分析仪检测系

统进行定期校准,既能够避免不必要的校准、又能够节省检验工作的时间和成本,保证生化检验项目检测

结果可靠性。

关键词 生物化学;仪器与设备;校准周期;日立H7180

中图分类号R446 文献标识码文章编号1007—9564(2015)02—0332一O3

D0I编码 1O.11723/mtgyyx 1007—9564 201502049

SCIENTIFIC METHOD OF ESTABLISHING CALlBRATING PERIOD OF HITACHI 7180 FULLY AUTO—

MATIC BIOCHEMICAL ANALYZER Wang Chunyu,Li Yuzhou,Zhang Guili.Clinical Laboratory,

Beidaihe Hospital。Qinhuangdao 0661 O0,China

Abstract Objective To establish calibrating period of the fully automatic biochemical analyzer Hitachi

7180 scientifically by using calibrating periodogram.Methods With the calbration and adjustment of the

biochemical analyzer,after the scaling calbration time 0,2,4,6,8,10,12,24 hours,the quality control

sera of the high,middle and low concentration levels separately were tested at one time respectively,then

ook the test one time every 24 hours,with the total amount of time for 30 days.The testing time could be

prolonged for the stable items,the testing should be finished after appearance of deviation.A1lhe test i—

ters were tested repeatedly four times to get the average data.The SPSS 19.0 statistical software was used

to process data,calculate CV and draw calibrating periodogram.The interval time after the ending of call

bration tillhe coefficient of variation of deposition(CV)greater than 1/6 CI IA'88 allowable deviation to

end the test was the calibration period for this item.Results The shortest calibration period was ALP with

onl14 hours;the calbration period for Calum(Ca)was 24 hours;the calbraton perod or TP,AI B,

CK,Fe was 4 days to 10 days;period for DBil。TBil,HDI 一C,GLU。Urea,CHO was 10 days to 20

days,25 days for I DH。over 30 days for TG de—ionization,Cr enzymic method。UA,AMY。AI T,

AST.Conclusion Scientific and reasonable establishment of the calibrating period of each test item can

help to avoid unnecessary calibration and save time as well as cost,and ensure the reliability of the testing

results of the chemical testing items.

Key words Biochemistry;Instrument and faciity;Calibrating period;Hitachi 7180